Questions and Answers

  • M
    michaelschwartzAug 9, 2025
    Why is SYSMEX Laboratory Equipment staining too pink?
    • O
      omurrayAug 10, 2025
      If the staining with your SYSMEX Laboratory Equipment is too pink, this may be due to an incorrect proportion of azure B – eosin Y, a buffer pH that is too low, or excessive washing in the buffer solution.
  • J
    Jeffrey WatersAug 15, 2025
    What causes a blue background in SYSMEX Laboratory Equipment staining?
    • I
      ikelleyAug 15, 2025
      A blue background in SYSMEX Laboratory Equipment staining may be due to inadequate fixation or prolonged storage before fixation, or blood collection into heparin as an anticoagulant.
  • R
    Robert JohnstonAug 21, 2025
    What causes neutrophil granules to appear pseudo-toxic in SYSMEX Laboratory Equipment?
    • R
      Randall VanceAug 21, 2025
      Neutrophil granules appearing pseudo-toxic in SYSMEX Laboratory Equipment could be caused by excess azure B.
  • A
    angela34Aug 27, 2025
    Why are neutrophil granules not stained in SYSMEX Laboratory Equipment?
    • C
      Carolyn JenkinsAug 28, 2025
      If neutrophil granules are not stained when using SYSMEX Laboratory Equipment, this may be due to insufficient azure B.
  • M
    Melissa MeyerSep 2, 2025
    Why are nuclei too blue with SYSMEX Laboratory Equipment?
    • J
      Jeremy TravisSep 2, 2025
      The nuclei appearing too blue when using SYSMEX Laboratory Equipment could be due to eosin concentration being too low.

Summary

Analyzer Components

Smear Production and Staining

Smear Production Steps

Explains the two-step process of smear production, detailing slide movement and the role of the smear unit.

Cellular Staining Principles

Cellular Component Colouration

Details cellular components and their corresponding colours based on stain binding and affinities.

Staining Troubleshooting Guide

Common Staining Issue Resolution

Identifies common staining problems and provides potential causes for troubleshooting.

SP-Series Quick Usage Guides

Analyzer Maintenance Procedures

Daily Maintenance Tasks

Details on performing daily maintenance, including shutdown procedures and checking the trap chamber.

Spreader Glass Cleaning

Step-by-step guide on how to clean the spreader glass using moistened gauze and deionized water.

Weekly Maintenance Tasks

Instructions for performing weekly maintenance, including Shutdown 2 and cleaning single cassettes.

Analyzer Start-Up Procedure

As-Required Maintenance Tasks

Running Patient Samples

Analyzer Status Monitoring

Reagent Replacement

Responding to Analyzer Alarms

Alarm Levels, Colours, and Meanings

Explains the different alarm levels, their associated colours, sounds, and their meaning for operator action.

Error List and Error Details

SYSMEX SP-10 Specifications

General IconGeneral
BrandSYSMEX
ModelSP-10
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals