Shutdown 2 (Weekly)
Ensure analyzer status is READY. From main menu select the [Shutdown] to open
the shutdown menu. Select the [Shutdown2 (Weekly)] to open confirmation screen. Ensure that all
cassette and methanol requirements are fulfilled.
• 8 Slide cassettes in supply table
• 700ml of methanol
• 2ml of CellClean in the sample rack in front of the aspiration position
Place 2mls of CellClean into a blank sample tube with cap on, Place tube in position 10 of sample
rack on left hand side of sampler unit and align with mark.
Press [OK]. The shutdown progress screen will
appear and once shutdown is complete analyzer
will turn off. Turn SP-series analyzer [OFF] at
mains power switch. Perform analyzer start-up
procedure (see page 22)
Cleaning Single Cassettes
Single slide cassettes should be clean on a regular basis to
prevent stain deposit build up. Remove cassettes from
analyzer and fill each cassette with methanol. Immediately
discard methanol and allow to air dry before use. Ensure there
is no residual methanol prior to use.
IMPORTANT: Do NOT leave methanol in cassettes for longer
than 10mins.
NOTE: Ensure cassettes are thoroughly dried before use.
Methanol within cassettes can cause haemolysis and outside the cassettes can cause interference
with transport of cassettes.