Audit logs
The SP-series analyzer has 3 audit logs
• Error History
• Parts history
• Reagent history
•
The ‘error history’ screen holds 1000 errors. To view error history from the main menu press
[Maint.], then [history} and [Error history].
The error history screen displays
The ‘parts history’ screen holds 100 entries. To view supply parts history from the main menu press
[Maint.], then [history} and [Parts history].
The ‘parts history’ screen displays the supply part
replaced along with the date and time of replacement.