TROUBLESHOOTING
Sysmex KX-21 Operator’s Manual -- Revised October 1998 7-7
3. TROUBLESHOOTING GUIDE
3.1 Pressure/Vacuum Errors
Error Message Meaning of Message
"Pressure/Vac Error"
0.35 [0.40 - 0.60]
ERR CODE:XXXXX.X.X
*
Help
*
0.5kg/cm2 Pressure Error
Adjust 0.5kg/cm2 Pressure.
Restarting to monitor the pressure[1]
1:End of Adj
0.5 kg/cm
2
is out of the monitor range in Ready
status.
Monitor range: 0.40 - 0.60 kg/cm
2
Cause 1) 0.5 kg/cm
2
pressure adjustment error
2) Pneumatic unit pressure error
3) Air leakage from pressure line
Action
1) Adjust Pressure to 0.5 kg/cm
2
.
The HELP screen displays pressure value in real time.
Adjust pressure referring to
Chapter 8, Section 2.3: Adjusting
Pressure to 0.5 kg/cm
2
.
2) Check Pressure Line
Check the pressure line for loosened nipple or tube, or breakage.
If any abnormality is noted, reconnect or replace.
After adjustment or check, press [1] key to return to the original
screen.
Analysis after error
occurrence
Not Ready until the error is resolved.