TROUBLESHOOTING
Sysmex KX-21 Operator’s Manual -- Revised October 1998 7-9
Error Message Meaning of Message
"Pressure/Vac Error"
ERR CODE:XXXXX.X.X
*
Help
*
0.5kg/cm2 Pressure Error at count
Turn OFF then ON the power.
3:Ana.Screen
0.5 kg/cm
2
pressure is out of the monitor range
during analysis.
Monitor range: 0.30 - 0.70
Cause 1) 0.5 kg/cm
2
pressure adjustment error
2) Pneumatic unit pressure error
3) Air leakage from pressure line
Action 1) Adjust Pressure to 0.5 kg/cm
2
Turn the power OFF, and turned it ON again. Then adjust pressure
referring to
Chapter 8, Section 2.3: Adjust 0.5 kg/cm
2
Pressure
.
2) Check Pressure Line
Check the pressure line for loosened nipple or tubing, or
breakage. If any abnormality is noted, reconnect or replace.
After adjustment or check, press [1] key to return to the original
screen.
Analysis after error
occurrence
The power needs to be turned OFF then ON again.
NOTE: Press [3] key to return to the Analysis screen. The status is not
ready for analysis, but the processing, etc. of stored data in Select
Menu can be performed.