EasyManua.ls Logo

SYSMEX KX-21 - Page 20

SYSMEX KX-21
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INTRODUCTION
Sysmex KX-21 Operator’s Manual -- Revised October 1998 1-3
3. OPTION UNITS
This instrument offers several option units to ensure its efficient operation. The options that
can be used with the KX-21 are:
Serial interface: Outputs analysis data and QC data to the host computer. Ordering
Information: OUTPUT UNIT NO. 1 KX-21 (Part No. 973-4531-1)
4. OUTLINE OF OPERATION
Two analysis modes are available with the KX-21: whole blood mode and pre-diluted mode.
Analysis procedures in these modes are listed below.
Whole blood mode Pre-diluted mode
Check before turnin
g
ON the power.
Turn ON the power.
Self-check
Back
g
round check
Read
y
Select whole blood mode. Select pre-diluted mode.
Set sample No. Prepare anal
y
sis samples in pre-diluted mode
(
1:26 dilution
)
.
Set sample No.
Set sample to the sample probe.
Press the start switch.
Execute anal
y
sis.
End anal
y
sis.
Read
y
Check after anal
y
sis.
Execute shutdown.
Turn off the power.
Table 1-4-1: Outline of Operation
The work by an operator is shown in shaded cells.
: "Ready" is displayed on the LCD screen. In this state, various operations
including analysis, settings, and data processing can be performed.
5. AUTOMATIC STOP FUNCTION OF PNEUMATIC UNIT
The KX-21, when left non-operating for 15 minutes, automatically stops the pneumatic unit.
This function saves power consumption and extends component service life. In addition,
with this function, analysis-ready status can be resumed faster than by turning on the power.
It is also possible to manually stop the pneumatic unit through the Select Menu screen.
Ready
Sysmex KX-21 Operator’s Manual -- Revised January 2000

Related product manuals