TROUBLESHOOTING
Sysmex KX-21 Operator’s Manual -- Revised October 1998 7-29
Error Message Meaning of Message
*
Scheduled Maint
*
1:Exec.Clean
Cycle No. after cleaning. 2567
Date last serviced 31/12/97
3:Cancel
Clean the Transducer.
It will take approx. 7 minutes.
It is time to clean the transducer.
One month has passed from the previous
cleaning.
2500 samples have been analyzed from the
previous cleaning.
Cause
Action
Press [1] key. Following the instructions on the screen, clean the
transducer (Refer to
Chapter 4, Section 5.2: Clean Transducer
).
If you perform cleaning later, press [3] key. This activates the usual
startup sequence and makes the system Ready.
The same message will appear the next time you start the unit; it will
be displayed at each startup until the cleaning is performed.
Analysis after error
occurrence
Ready for analysis
NOTE: When the transducer rinse sequence has been completed, the cycle
counter is automatically reset.