NB/NHB Series Technical Manual
- 23 -
Remove the load and restart
scale again.
Err 18
Pre-tare error
Check the pre-tare value
Err19
Initialize zero error
Calibration the scale.
Err P
Printer error
Check the printer settings or
connections
Err l
Approval setting error
Check the PCB jumper
settings. Must be connect
jumper pin to K1 (BW series)
--ol--
Over range
Remove the load.
Re calibrate
--lo--
Underload
Minus weight, check the
platform and restart or
calibrate.
Fai l h /
fai l l
/
fai l
Calibration Error
Check the test weights & Re
calibrate
Ba lo /
lo ba
Battery low
Re charge battery, check the
voltages.