EasyManua.ls Logo

TabTimer Careousel MK3/11 - Page 9

TabTimer Careousel MK3/11
12 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
Baeries:
Baeries (4 x AA Alkaline) are supplied with the dispenser when new.
Baeries will last in normal use for up to 12 months.
Baery life can be checked using the following procedure:
1. Insert baeries in the dispenser.
2. Immediately watch the LCD display for the indicaon BATLEV.
3. At full charge a reading of between 1550 - 1600 (or more) will be shown.
4. At a reading of 1325 the baery warning indicator ‘’LOWBATT’’ will iniate.
5. At a reading of 1275 the unit will not operate, will blink ERROR and sound alarm
sound 04 for 2 seconds, silence for 28 seconds, repeated. This is dierent to the
standard alarm which sounds for 2 seconds repeang every 2 seconds.
NB: 1. Rechargeable baeries should not be used as they have an insucient charge to
reliably operate the unit.
2. When the baery charge is reduced to a reading of 1325 the display will indicate
LOWBAT. In normal operaon the LOWBAT warning will be acve for 1-2 months
depending on how many daily doses are set. This gives early warning that a baery
change is required.
Remember when changing baeries to insert them the correct way around (see gure 3. – p.3).
Removing Baeries: If the baeries are removed and replaced immediately the alarm
sengs are retained; only the clock would need to be reset if the baeries are not replaced
immediately.
Cleaning:
Dispenser - clean external surfaces of the dispenser body and lid with a disinfectant
spray and / or wipe with a damp cloth.
Internal tray - the tray may be washed in warm water or in a dishwasher (max. 35C).
Transport: If for any reason the dispenser is despatched by post, courier or other carrier, it is
important that all baeries are removed from the dispenser.
Important - Reposioning the Tray. To avoid damaging the mechanism, the medicine tray
must always be reposioned by liing it out and then lowering directly into the case in the
correct posion. Do not force the tray to rotate.
TabTimer cannot accept responsibility for faults or any other consequences arising from
misuse e.g. dropping, water damage, inseron of foreign objects, forced turning of the pill
tray or forced opening of the dispenser.