Medix 24
15
RFID
RFID reader supports RFID card tapping (optional)
Optical Disc Drive
8X DVD RW SATA interface (optional)
Adapter
Input: 100-240V~1.5A 50/60Hz
Output: 19V/4.7A, max. 90W
(Medical Adapter:
FSP FSP090-RBBM1)
IP Rating
IP65 front panel , IPX1 Top
Thermal Solution
Fanless housing
VESA Wall Mount
75x75 & 100x100mm VESA mounting,
Compatible with ARS5 Arm Head (Quantity : 4 pcs)
The VESA mounting holes at the back of the terminal are
provided with M4-type blind fasteners to fix the VESA mounting
plate. Depending on the VESA plate thickness (T) and the
thickness of possible washers (W), a different screw length (L)
should be selected.
Please respect the following rule to select an appropriate
screw length:
L
min
= T + W + 5 mm
Lmax = T + W + 10 mm
Arm selection suggestions:
Includes VESA standards with hole pattern bracket.
The designed using aluminium alloy and plastic materials.
Max.Load-bearing: 10 kg or more
Temperature
Operational: 0°C~40°C (32°F~104°F)
Storage: -10°C~50°C (14°F~122°F)
Transport: -10°C~+50°C (14°F~122°F)
Humidity
Operational: 20%~80% @40°C non-condensing
Transport: 15%~95%
Atmospheric Pressure
Range
Operational: 650 to 1060 hPa
Transport: 375 mmHg to 800 mmHg (500 hPa to 1060 hPa)
Operation Altitude
4000m
Operating System
Windows 10 (optional)
Safety and Certifications
UL 60601-1 3.1 Edition
FCC Class B
EN 60601-1-2 4th Edition
CE
IEC 60601-1-2 3.1 Edition
Energy Star 8.0
Other Features
Reading light
Camera privacy latch
Description