EasyManua.ls Logo

Techno Control SE137E Series - TEST and CALIBRATION

Techno Control SE137E Series
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IST-2137.EC03.01/E Istruzione / User’s Manual / Manuel d’utilisation Pag. 11/20
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) Tel. +39 02 26922890 - Fax +39 02 2133734
paints. However, in case of products containing substances in great quantity, these could interfere with the Sensor
and cause false alarms. We recommend ventilating the room when products like these are used. Alcohols, Hydro-
gen and volatile organic solvents in great quantity cause false alarms, and then the cell needs one or more hours in
clean air to recover the normal sensitivity conditions.
TEST and CALIBRATION
PAY ATTENTION: This procedure has to be made with extreme attention and by authorized and trained people;
because starting this procedure it will increase mA output causing the activation of connected alarm devices.
Operation Check, Zero Adjust, and Calibration: are different code protected functions. To access these func-
tions is necessary to insert the relevant Code" through the keys F1s and F2. To have the key pressure recog-
nized, hold pressing it for around a second (until the green Led doesn't switch off for a moment). Then the next key can be
pressed. In case of error all it takes is waiting around 10 seconds and the sequence is automatically erased.
The Calibration Check is a function code free.
Calibration Kit, Sample Gas Bottles (for Calibration Check and Calibration) the mixture to be used depending on the
model installed is shown in the table below:
MODEL
Sample gas bottle
Our Code KIT Code
S
E137
EA
-
S
E137
EA
-
H
100 ppm Ammonia (NH
3
) in Nitrogen Available on request TC014
S
E137
EC
-
S
-
S
E137
EC
-
H
300 ppm Carbon monoxide (CO) in Nitrogen BO210 TC011/TC014
S
E137
ES
10 ppm Sulphur dioxide (SO
2
) in Nitrogen Available on request TC014
S
E137
EH
50 ppm Hydrogen sulphide (H
2
S) in Nitrogen Available on request TC014
S
E137
EHCN
10 ppm Hydrogen cyanide (HCN) in Nitrogen Available on request TC014
S
E137
EN
100 ppm Nitric oxide (NO) in Nitrogen Available on request TC014
S
E137
EN2
30ppm Nitrogen dioxide (NO
2
) in air Available on request TC014
S
E137
EHC
L
10ppm Hydrogen cloride (HCL) in Nitrogen Available on request TC014
S
E137
EC
L
10ppm Nitrogen dioxide (NO
2
) in air Available on request TC014
It is possible to use either the disposable cylinders with adjust valve or the high pressure ones with reduction
gear. Some cylinders have a fixed flow reducers, make the necessary value in Table, column 3, page 11. Is also nec-
essary the Tecnocontrol calibration kit model TC014 (Inox). To ensure that the gas concentration reaches sen-
sor, use steel reduction gear and Teflon tubes or other compatible materials which do not disperse gas.
“INSTRUMENT OPERATION CHECK” (Check Code: F2, F2, F1, F1): this function allows to effect a functional test
of the detector. After having inserted the "Check Code", all Led are switched off and all the relays are deactivated.
Then they will switch on in sequence, the Led, from the yellow up to the red. To the lighting of the various Led it
corresponds the activation of the relevant relays (relay "FAULT" with the yellow Led, relay "ALARM1" with the red Led, relay
"ALARM2" with the red Led, relay "ALARM3" with the red Led). At the end all the Led will remain lighted for around 5
seconds, then the detector returns at the conditions of normal operation. It is advisable to perform this operation
every 12 months according to the use. Note: this function is not working if the and/or the and/or the 3° red
Led have already turned on.
“ZERO ADJUST” (Zero Code: F2, F1, F1, F2): this function is to adjust the Zero sensor and can be done in clean
air only (environment without the presence of toxic gas or other pollutants). Consider that the SE137E series have a sophisti-
cated follower of zero, that every hour will reset the sensor, if of course there is the presence of gas. If necessary,
perform this operation after installation or after changing the Cartridgeor every 6-12 months depending on envi-
ronmental conditions. After entering the "Zero Code" as confirmation of the operation, there will be a flashing 1
st
red
Led and output becomes 4.0 mA.
Note: This function is not working if mA output is higher than the value shown in the table below, depending on the
model, or if the 2
nd
and/or 3th red Led are already ON. In this case, please recalibrate the unit and/or replacing the
Cartridge”. In case the 1
st
red Led will be already ON, only instead of flashing it will switch off for confirmation.
SE137
EA
SE137EA-H
SE137
EC
-
S
SE137EC-H
SE137ES SE137EH SE137EHCN SE137EN SE137EN2 SE137EHCL SE137ECL
4,8 mA
15ppm NH
3
4,8 mA
15ppm CO
6,4 mA
3ppm SO
2
4,8 mA
5ppm H
2
S
8 mA
2,5ppm HCN
5,6 mA
10ppm NO
6,6 mA
5ppm NO
2
8 mA
2,5ppm HCL
4,
8 mA
0,5ppm HCL
“CALIBRATION” (Calibration Code: F2, F2, F2, F1, F2, F1): this function allows completely recalibrate the sensor
using the gas mixture indicated above.
Warning: to guarantee that no errors of elaboration happen, the possibility exists that during the Calibration the yel-
low Led switch off every 8 seconds, in this case interrupt the procedure, switch off and switch on the instrument and
repeat the Calibration. If condition persists it will be necessary to send the detector to the supplier for the reparation.
Important note: during Calibration routine the mA output indicates 0mA.
The Calibrationcan be done in clean air only (environment without the presence of toxic or other polluting gas). With the
keys perform the Calibration Code". Wait until the yellow and green Led switch on fix and the 1
st
red Led starts to
flash. Insert the TC011 on the sensor holder and adjust the influx of the gas the way that the flow meter reaches
the value indicates in table below in column 3, depending on the model (see also Fig.3).
Wait around 3-5 minutes, until the 3
rd
red Led lights (and while it is ON), press the key F2 and hold it pressed until
the 1
st
and 3
rd
red Led are switched off for at least 2 seconds (but if the 1
st
red Led continues to blink more than 5-6
minutes means that the gas flow is not constant or unsatisfactory, or the concentration is not as required, stop the
operation, turn off the instrument and recalibrate). Then, close the gas cylinder and remove TC011. At this point we
can have two possibilities:

Table of Contents

Related product manuals