PROVASET 2P code 2P2B00 - 2P6B00 - Instruction manual
rev. 14/02/2013
page 10
! CAUTION: The protective earth terminal is connected to the tester chassis.
Check the good state of the earth connection while installing the
equipment and later periodically according to standards.
! CAUTION: Disconnect the power supply wire before replacing the fuses.
9. Compressed air inlet connector:
Quick connector for tube 6x4 mm.
Inlet max. pressure: 10 bar.
% WARNING: The supply pressure must be kept constant to allow a correct operation
of the pressure reducer; pressure changes can modify the test pressure.
! CAUTION: Use only dry air, without condensate, filtered, not lubricated and free
from explosive gases.
10. Label:
It indicates the producer data, the tester exact code, its serial number and the data
relating to the necessary electrical voltage supply.