PROVASET T3LPF REV.20210610
47
The PR_MAX-
grammed as test pressure.
The minimum test pressure is programmable from PR = 0, but actually
/ Pressure PR
= TIME) or when the PR test pres-
sure
PR = PRESS.) is reached.
This mode is used to save time and move immediately to the next step (settling time
) as soon as the PR test pressure is reached.
, to reach
PR in one or more steps.
Field 0: 255 steps; 0=no ramp.
end part.
The parameter indicates the percentage of rising/falling time for each step.
Dimension: 0: 100%.
Settling phase
T2
Settling time, range 0.0: 3600.0
Waiting time to stabilise the pressure in the test circuit before the measurement
phase
PR%MAX: maximum % limit of the test pressure, range 0.0: 100.0 %.
The test ends with a result
if the limit is exceeded.
For leak tests:
• maximum test pressure P < PR + PR * (PR%MAX/100);
• limit applied during all phases
;
• limit not applied if
PR = 0.
PR%MIN: minimum % limit of the test pressure, range 0.0: 100.0 %.
The test ends with a result
if the limit is exceeded.
For leak tests:
• limit applied during all phases
;
• limit not applied if
PR = 0.