Betriebsanleitung/Operating manual
E3905 / P3394
32 BD_BE 763 b www.tecsis.de
Conf
Set Configuration: General configurations
Sea-level correction: on or off
(Alti)
-2000..9999
(only if SL is on)
Altitude: Input of altitude above sea level in [m]
rAtE**
Slo
Slow measuring rate
(4Hz filtered, low power consumption)
Fast: measuring rate, filtered (>100Hz)
P.dEt
Peak detection :fast measuring rate,
unfiltered (>100 Hz)
P.OFF
Auto Power-Off time in minutes
Auto Power-Off deactivated
Base address of interface
AL.
Set Alarm: Settings of alarm function
AL.
[1,2,DIF]
Alarm on, with horn-sound
Alarm on, without horn-sound
(not if AL.oFF)
Min-alarm-limit
( AL. oFF)
Max-alarm-limit
LOGG
Set Logger: Settings of logger function
Func
Cyclic: logger function „ cyclic logger“
Store: logger function, „individual value logger“
(CYCL) 1..3600
(only if Func CYCL is set)
Cycle time of cyclic logger in [seconds]
(Lo.Po) ON / OFF
(only if cyclic logger and slow measuring rate is
set)
Low-Power-Logger with very low power
CLOC
Set Clock: Settings of real time clock
Clock: Setting of time: hour : minutes
Year: Setting of time: year
Date: Setting of time: day : month
Please note: If the logger memory contains data already, the menus/parameters marked with (**)
can not be invoked! If these should be altered the logger memory has to be cleared
before (p.r.t. chapter 6)!