EasyManua.ls Logo

TEKNA DPG601 - Page 19

TEKNA DPG601
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
901
16
6
0,25
1000
250
25
2,50
0,25
14
7
0,29
1160
290
29
2,90
0,29
12
8
0,33
1320
330
33
3,30
0,33
902
10
10
0,42
1680
420
42
4,20
0,42
6
12
0,5
2000
500
50
5,00
0,50
2
15
0,63
2520
630
63
6,30
0,63
903
5
25
1,04
4160
1040
104
10,40
1,04
4
40
1,67
6680
1670
167
16,70
1,67
3
50
2,08
8320
2080
208
20,80
2,08
904
2
60
2,5
10000
2500
250
25,00
2,50
1
80
3,33
13320
3330
333
33,30
3,33
0
100
4,17
16680
4170
417
41,70
4,17
Adj
usting of the working frequency
By adjusting the maximum working frequency (400 strokes/min) we can set up the speed with which the “n” strokes are given
after the arrive of the signal from the water meter pulse sender.
Example: Setting up n=40, to every impulse from outside the pump will dose 40 strokes.
At the maximum working frequency (400 strokes/min) the pump will employ 6 second to dose 40 strokes; if
between two successive signals they passed, as an example, 50 second ones, we could have excess dosaga; if we
wished to have a dosage for at least 30 seconds, we have to set the pump with a working frequency of 80
strokes/min.
Warning:
please test if the pump is fast enough, so as to finish dosing before the arrival of the next signal from the water
meter; if not the memory function alarm will be activated.
Example mode n : 1
I
have to dose 1.000 ppm in one lead with a pressure of 2 bar, having:
- water meter mod. CB4 (4 imp/l)
- dosing pump mod. Tekna DPG 904 (60 l/h a 2 bar)
In the table on the previous page we see that the model 904 and the water meter CB4, with n=1 doses 10.000 ppm
The value of concentration of the dosage in table with n=1 is
too high regarding what it is necessary.
In such case I will have to program the DPG pump in mode n : 1.
Setting up n=10, I obtain the 1.000 ppm wished.
Adjusting of the working frequency
Adjusting the maximum working frequency (400 strokes/min) we can set up the speed at which the “n” strokes are given after
the arrival of the signal from the water meter.
Wa
rning:
please test if the pump is fast enought , so as to finish dosing before the arrival of the next signal from the water
meter; if not the memory function alarm will be activated.
Example:
A water meter pulse sender mod. 1 stroke/L with a flow rate of 15 mc/h, will give 250 strokes/min; setting the pump
with "n"=5, the pump will dose 50 strokes/min (50 = 250/5).