EasyManuals Logo
Home>Tektronix>Test Equipment>475

Tektronix 475 User Manual

Tektronix 475
251 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
Calibration475
b. Connect a 50 mV signal from the Standard Amplitude
Calibrator to the CH 1 input.
c. ADJUSTCH 1 Position Centering, R176, so the top
and bottom of the display can be positioned past the center
horizontal line.
d. Repeat the above for CH 2 adjust R276.
16. Adjust CH 2 Invert Balance (R215) and Check
Operation
17. Adjust CH 1 Gain
a. CH 1 VOLTS/DIV to 5 mV.
b. Connect 20 mV signal from Standard Amplitude
Calibrator to CH 1 input.
c. ADJUST-CH 1 5 mV Gain, R165, for 200 mV push-
pull signal between preamp sides of the delay line.
d. CH 1 VO LTS/DIV to 2 mV.
e. ADJUSTCH 1 2 mV Gain, R125, for 100 mV push-
pull signal between preamp sides of the delay line.
f. ADJUSTR449 for 5 divisions of deflection.
g. Check all attenuator ranges for proper deflection,
within 3%.
h. Check Variable Range: CH 1 VAR control must
reduce a 5 division signal to less than 2 divisions.
18. Adjust CH 2 Gain
a. CH 2 VOLTS/DIV to 5 mV.
b. Connect 20 mV signal from Standard Amplitude
Calibrator to CH 2 input.
c. ADJUSTCH 2 5 mV Gain, R265, for 4 divisions
deflection, within 3%.
d. CH 2 VOLTS/DIV to 2 mV; test signal to 10 mV.
5-16
e. ADJUSTCH 2 2 mV Gain, R225, for 5 divisions
deflection, within 3%.
f. Check all attenuator ranges for proper deflection,
within 3%.
g. Check Variable Range: CH 2 VAR control must
reduce a 5 division signal to less than 2 divisions.
19. Check ADD Operation
a. Both VOLTS/DIV to 5 mV,
both AC-GND-DC to DC.
b. Connect a 10 mV signal from the Standard Amplitude
Calibrator to both CH inputs.
c. VERT MODE to ADD.
d. Check for 4 divisions deflection, within 3%.
20. Check Compression and Expansion
a. Position a 2 division signal to top and bottom of
graticule.
b. Check for no more than 0.1 division of compression
or expansion.
21. Check ALT Operation
22. Check CHOP Operation
a. T IM E /D IV to 0.2 jus,
VERT MODE to CHOP.
b. Check for blanking of transients at normal INTEN
SITY level.
c. Check duration of each cycle is about 5 divisions.
23. Adjust Vertical Output High-Frequency
Response
®
a. VERT MODE to CH 1.

Other manuals for Tektronix 475

Questions and Answers:

Tektronix 475 Specifications

General IconGeneral
BrandTektronix
Model475
CategoryTest Equipment
LanguageEnglish

Related product manuals