EasyManuals Logo
Home>Tektronix>Test Equipment>6 Series

Tektronix 6 Series Declassification And Security Instructions

Tektronix 6 Series
17 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #13 background imageLoading...
Page #13 background image
Media and data export devices
The following table lists the data export devices in the instrument.
Table 3: Media and Data export devices
Type Function
Method of
modification
Data input
method Location
User
accessible To disable
USB Host
port
(supports
removable
USB flash
drive)
User storage of
reference
waveforms, screen
images, and
instrument setups,
and installation of
firmware updates
Directly User
writeable
Three USB
host ports on
front panel of
the
instrument;
four USB
host ports on
the back
panel of the
instrument
No
USB Host ports can be
disabled for use with
USB Storage devices
behind password
control.
Note: This
option is
available if
Option 6-SEC
is installed.
USB Device
port
Supports remote
control and data
transfer to a PC
Directly Remote
control using
USBTMC
USB Device
port on back
of the
instrument
Yes
The USB Device port
can be disabled by
selecting Utility > I/O >
USB DevicePort, and
setting USB Device port
to Off.
Note: This
option is
available if
Option 6-SEC
is installed.
Ethernet Transfer data and
remote control of
instrument.
Directly Remote
Control using
LXI or Socket
Server
Ethernet port
on back of
instrument
No Ethernet port can be
disabled behind
password control
Note: This
option is
available if
Option 6-SEC
is installed.
Option 6-SEC for a secure instrument
Option 6-SEC provides the highest level of instrument security for 6 Series B MSO products. Option 6-SEC features include:
Password protection to enable/disable external communication ports
Password protection to enable/disable firmware upgrades or downgrades
There is also a special BIOS installed that includes a default password ("Tektronix"). Additionally when the CMOS reset
is pressed the BIOS password is reset to "Tektronix" instead of being removed.
Note: Option 6-SEC must be ordered at the same time as ordering an instrument.
Clear and sanitize procedure
6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions 7

Other manuals for Tektronix 6 Series