EasyManuals Logo
Home>Tektronix>Multimeter>DMM912

Tektronix DMM912 Technical Reference

Tektronix DMM912
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #250 background imageLoading...
Page #250 background image
PS2520 Series Performance Verification
Handheld and Benchtop Instruments Basic Service
31
2. Press SHIFT
PARA/INDEP to configure OUTPUT 1 and OUTPUT 2 for
parallel operation. Verify that the “PARA” indicator lights up on the display.
3. Set the DMM to measure 5 amperes DC.
4. Set up the power supply as follows:
VOLTS SET 36 V
CURRENT SET 3.0 A
OVP SET 38.5 V
5. Press OUTPUT ON/OFF. Verify that the “OUT” indicator lights up on the
display.
6. Adjust the output of the Variac from 108 to 132 VAC (120 V range) or 198
to 242 VAC (220 V range). Verify that the DMM current variation is 0.006
A over the adjustment range.
7. Press OUTPUT ON/OFF. Verify that the “OUT” indicator turns off.
8. Press SHIFT
PARA/INDEP to reconfigure the outputs for independent
operation. Verify that the “PARA ”indicator turns off.
Output 3. Use the following steps to check the OUTPUT 3 accuracy.
1. Press SHIFT
OUT 3; verify that the “3” indicator lights up on the display.
2. Ensure that the power supply output is disabled. Connect the DMM and
2.2 W 50 W resistor to the front panel OUTPUT 3 terminals. See Figure 6
for details.
3. Set up the power supply as follows:
VOLTS SET 6 V
CURRENT SET 3.0 A
OVP SET 7 V
4. Repeat steps 6 through 8 of the Outputs 1 and 2 procedure above.
This concludes the PS2520 and PS2520G performance verification procedures.

Table of Contents

Other manuals for Tektronix DMM912

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DMM912 and is the answer not in the manual?

Tektronix DMM912 Specifications

General IconGeneral
BrandTektronix
ModelDMM912
CategoryMultimeter
LanguageEnglish

Related product manuals