DMM91X Series Adjustment Procedures
32
Handheld and Benchtop Instruments Basic Service
Adjustment Procedures
This section contains procedures to adjust DMM912, DMM914, and DMM916
multimeters. Perform these procedures once a year or if the Performance
Verification procedure indicates the need for calibration.
In this section you will find the following information:
A list of adjustments
A list of test equipment needed to make the adjustments
Instructions on how to prepare the instrument for adjustment
Step-by-step adjustment procedures
The procedures in this section do not verify performance. To confirm that your
multimeter meets factory specifications, perform the procedures in the Perfor-
mance Verification section.
List of Adjustments
Use the adjustments listed in Table 9 to return DMM91X Series multimeters to
factory calibration.
Table 9: DMM91X Series Adjustments
Adjustments Part 1
DC Volts
AC Volts
Capacitance
Temperature (DMM914 and DMM916)
DC Milliamperes
DC Amperes
Adjustments Part 2
AC Response

Table of Contents