PS2520 Series Adjustment Procedures
66
Handheld and Benchtop Instruments Basic Service
cL41 Parallel Output: Current Full Scale Calibration. To calibrate the parallel output
full-scale current, perform the following steps.
1. Verify that the “PARA” indicator lights up on the display.
2. Enter the DMM current reading on the power supply keypad and press (
).
The readout will return to normal when the calibration is complete.