EasyManua.ls Logo

Tektronix DMM916

Tektronix DMM916
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TAS 200 Series Performance Verification
12
Handheld and Benchtop Instruments Basic Service
NOTE. At lower VOLTS/DIV settings, you might need to set TRIGGER COU-
PLING to HF REJ to minimize noise or double triggering.
Table 14: DC Coupled Bandwidth
Volts/Div 50 kHz
Reference Amplitude
Bandwidth
Minimum Amplitude
TAS 220
Bandwidth
TAS 250
Bandwidth
1 mV 6 divisions 4.2 divisions 10 MHz 15 MHz
2 mV 6 divisions 4.2 divisions 10 MHz 15 MHz
5 mV 6 divisions 4.2 divisions 20 MHz 50 MHz
10 mV 6 divisions 4.2 divisions 20 MHz 50 MHz
20 mV 6 divisions 4.2 divisions 20 MHz 50 MHz
50 mV 6 divisions 4.2 divisions 20 MHz 50 MHz
100 mV 6 divisions 4.2 divisions 20 MHz 50 MHz
200 mV 6 divisions 4.2 divisions 20 MHz 50 MHz
500 mV 6 divisions 4.2 divisions 20 MHz 50 MHz
1 V 4 divisions 2.8 divisions 20 MHz 50 MHz
4. Repeat steps 1 through 3 for CH 2.
To check DC balance accuracy, perform the following steps.
1. Set up the oscilloscope as follows:
VERTICAL MODE CH1 (CH2)
CH 1 (CH 2) GND In
CH 1 (CH 2) VARIABLE CAL
2. Verify that the oscilloscope trace shift is less than ±0.5 divisions when
rotating the CH 1 (CH 2) VOLTS/DIV control between the 5 mV and 5 V
settings.
3. Verify that the oscilloscope trace shift is less than two divisions when the
CH 1 (CH 2) VOLTS/DIV control is changed from the 1 mV to the 2 mV
setting.
4. Repeat steps 1 through 3 above for CH 2.
DC Balance

Table of Contents

Other manuals for Tektronix DMM916

Related product manuals