EasyManuals Logo

Tektronix DMM916 Technical Reference

Tektronix DMM916
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #313 background imageLoading...
Page #313 background image
TAS 200 Series Adjustment Procedures
Handheld and Benchtop Instruments Basic Service
25
Following the warm-up period, preset the oscilloscope to the settings listed
below.
Table 20: Oscilloscope Initial Settings
Control Setting
INTENSITY Visible display
CURSOR ON/OFF Off
VERTICAL
CH 1 (CH 2) POSITION Midrange
MODE CH1 (CH2)
CH 1 (CH 2) VOLTS/DIV 10 mV
CH 1 (CH 2) VOLTS/DIV VARIABLE CAL
CH 1 (CH 2) GND In
HORIZONTAL
POSITION Midrange
X10 MAG Out
X-Y Out
SWP UNCAL Out
TRIGGER
LEVEL Midrange
MODE AUTO
COUPLING AC
SOURCE CH 1 (CH 2)
SLOPE Rising (push
switch out)
Initial Settings

Table of Contents

Other manuals for Tektronix DMM916

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DMM916 and is the answer not in the manual?

Tektronix DMM916 Specifications

General IconGeneral
BrandTektronix
ModelDMM916
CategoryMultimeter
LanguageEnglish

Related product manuals