Commands Listed in Alphabetical Order
Syntax
TRIGger:A:BUS
:B<x>:I2C:ADDRess:MODe {ADDR7|ADDR10}
TRIGger:A:BUS:B<x>:I2C:ADD Ress:MODe?
Arguments
ADDR7 specifies the 7-bit I
2
C address mode.
ADDR10 specifies the 10-bit I
2
Caddressmode.
Examples
TRIGGER:A:BUS:B1:I2C:ADDRE SS:MODE ADDR10 sets the I2C address mode
to 10-bit.
TRIGger:A:BUS:B<x>:I2C:ADDRess:TYPe
Sets or returns the I
2
C address type. The only supported address type is USER.
Applies to bus <x>, wh
ere x is the bus number.
Conditions
This command requires a DPO2EMBD application module.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:I2C:ADD Ress:TYPe
{GENeralcall|STARtbyte|HSm ode|EEPROM|USER}
TRIGger:A:BUS:B<x>:I2C:ADD Ress:TYPe?
Arguments
GENeralcall specifies a general call address.
STARtbyte specifies a start byte address.
HSmode specifi es a high-speed mode address.
EEPROM specifies an EEPROM address.
USER specifies a user address.
TRIGger:A:BUS:B<x>:I2C:ADDRess:VALue
Sets or returns the binary address string used for the I
2
C trigger if the trigger
condition is ADDRESS or ADDRANDDATA. Applies to bus <x>, where x is the
bus number.
Conditions
This command requires a DPO2EMBD application module.
Group
Trigger
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-301