Commands Listed in Alphabetical Order
Arguments
AC sets channel
<x> to AC coupling.
DC sets channel <x> to DC coupling.
GND sets channel<x> to ground. Only a flat, ground-level waveform will be
displayed.
Examples
CH2:COUPLING GND sets channel 2 to ground.
CH3:COUPling? might return :CH3:COUPling DC indicating that channel 3
is set to DC coupling.
CH<x>:DESKew
Sets or returns the deskew time for channel <x>, where x is the channel number.
You can adjust the deskew time to add an independent, channel-based delay time
to the delay (set by the horizontal position control and common to all channels)
from the common trigger point to first sample taken for each channel. This lets
you compensate individual channels for different delays introduced by their
individual input hook ups.
Group
Vertical
Syntax
CH<x>:DESKew <NR3>
CH<x>:DESKew?
Arguments
<NR3> is the deskew time for channel <x>, ranging from -100 ns to +100 ns
with a resolution of 1 ns.
Examples
CH4:DESKew 5.0E-9 sets the deskew time for channel 4 to 5 ns.
CH2:DESKew? might return :CH2:DESKEW 2.0000E-09 indicating that the
deskew time for channel 2 is set to 2 ns.
CH<x>:IMPedance
Sets or returns the input impedance of channel <x>, where x is the channel
number. This command is provided for compatibilty.
Group
Vertical
2-100 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual