Commands Listed in Alphabetical Order
NOVershoot (ne
gative overshoot) nds the negative overshoot value over the
entire waveform or gated region. This measurement is available only on DPO
models.
Negative Overshoot = ((Low Minimum) / Amplitude) × 100%)
NPULSECount is the count of negative pulses.
NWIdth (negative width) measurement is the distance (time) between the middle
reference (default = 50%) amplitude points of a negative pulse. The measurement
is made on the rst pulse in the waveform or gated region.
PDUty (positive duty cycle) is the ratio of the positive pulse width to the signal
period, expressed as a percentage. It is measured on the rst cycle in the
waveform or gated region.
Positive Duty Cycle = ((Positive Width) / Period) × 100%
PEDGECount is the count of positive edges.
PERIod is the time required to complete the rst cycle in a waveform or gated
region. Period is the recipr oc al of frequency and is measured in seconds.
PHAse measures the phase difference (amount of time a waveform leads or lags
the reference waveform) between two waveforms. The measurement is made
between the middle reference points of the two waveforms and is expressed in
degrees, where 360° represents one waveform cycle.
PK2Pk (peak-to-peak) nds the absolute difference between the maximum and
minimum amplitude in the entire waveform or gated region. This measurement is
available only o n DPO models.
POVershoot is the positive overshoot value over the entire waveform or gated
region. This measurement is available only on DPO models.
Positive Overshoot = ((Maximum High) / Amplitude) ×100%
PPULSECount is the count of positive pulses.
PWIdth (positive width) is the distance (time) between the middle reference
(default = 50%) amplitude points of a positive pulse. The m easurement is made
on the rst pulse in the waveform or gated region.
RISe timing measurement nds the rise time of the waveform. The rise time is
the time it takes for the leading edge of the rst pulse encountered to rise from a
low reference value (default is 10%) to a high reference value (default is 90%).
This measurement is available only on DPO models.
RMS amplitude measurement nds the true Root Mean Square voltage in the entire
waveform or gated region. This measurement is available only on DPO models.
Examples
MEASUREMENT:MEAS2:TYPE FREQUENCY denes measurement 2 as a
measurement of the frequency of a waveform.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-205

Other manuals for Tektronix DPO2024

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2024 and is the answer not in the manual?

Tektronix DPO2024 Specifications

General IconGeneral
BrandTektronix
ModelDPO2024
CategoryTest Equipment
LanguageEnglish

Related product manuals