MEASUrement:
METHod
Sets or returns the method used to calculate the 0% and 1 00% reference level.
Group
Measurement
Syntax
MEASUrement:METHod {Auto|HIStogram| MINMax}
MEASUrement:METHod?
Related Commands
MEASUrement:REFLevel:PERCent:HIGH, MEASUrement:REFLevel:PERCent:
LOW, MEASUrement:REFLevel:PERCent:MID, MEASUrement:REFLevel:
PERCent:MID2
Arguments
Auto selects the best method for each data set.
HIStogram sets the high and low waveform levels statistically using a histogram
algorithm.
MINMax uses the highest and lowest values of the waveform record. This selection
is best for examining waveforms with no large, flat portions of a common value,
such as sine waves and triangle waves.
Examples
MEASUREMENT:METHOD? might return :MEASUR EMENT:METHOD MINMAX
indicating that the reference levels are set to MIN and MAX.
MEASUrement:REFLevel? (Query Only)
Returns the current reference level parameters.
Group
Measurement
Syntax
MEASUrement:REFLevel?
Examples
MEASUREMENT:REFLEVEL? might return these reference
level settings
:MEASUREMENT:REFLEVEL:METHOD
PERCENT;ABSOLUTE:HIGH 0.0000;LOW 0.0000;MID1 0.0000 ;MID 2
0.0000;:MEASUREMENT:REFLEV EL:PERCENT:HIGH 90.0000;LOW
10.0000;MID1 50.0000 ;MID2 50.0000
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-207

Other manuals for Tektronix DPO2024B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2024B and is the answer not in the manual?

Tektronix DPO2024B Specifications

General IconGeneral
BrandTektronix
ModelDPO2024B
CategoryTest Equipment
LanguageEnglish

Related product manuals