Commands Listed in Alphabetical Order
MEASUREMENT:M
EAS1:TYPE?
might return :MEASUREMENT:MEAS 1:TYPE RMS
indicating that measurement 1 is defined to measure the RMS value of a waveform.
MEASUrement:MEAS<x>:UNIts? (Query Only)
Returns the units associated with the specified measurement. The measurement
slots are specified by <x>, which ranges from 1 through 4.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:UNIts ?
Related Commands
MEASUrement:MEAS<x>:TYPe
Examples
MEASUREMENT:MEAS1:UNITS? might return :MEASUREMENT:MEAS1:UNIts %
indicating units for measurement 1 are set to percent.
MEASUrement:MEAS<x>:VALue? (Query Only)
Returns a calculate value for the measurement specified by <x>, whic h ranges
from 1 through 4.
NOTE. This is the same value as displayed on-screen. If measur
ement s tatistics
are enabled, a new value is calculated with every waveform. In addition, this
value is updated approximately every 1/3 second. If you are acquiring a long
acquisition record, the oscilloscope may take longer to update.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:VALue ?
Related Commands
MEASUrement:MEAS<x>:UNIts?, *ESR?, ALLEv?
Examples
MEASUREMENT:MEAS1:VALUE? might return : MEASUREMENT:MEAS1:VALue
2.8740E-06
. If the measurement has an error or warning as sociated with it,
then an item is added to
the error queue. The error can be checked for with the
*ESR? and ALLEv? commands.
2-206 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual