DC voltage measurement accuracy (averaged), >20 GHz models
Instrument performance test Minimum Incoming Outgoing Maximum
Ch1 6.25 mV Vert scale setting, 0 Div position setting
+3.5 V termination
+3.4 V offset
+2.4 V offset
+1.5 V termination
+2.5 V offset
+1.5 V offset
+0.5 V offset
0 V termination
0 V offset
-1 V offset
+1 V offset
-3.5 V termination
-3.4 V offset
-2.4 V offset
-1.5 V termination
-2.5 V offset
-1.5 V offset
-0.5 V offset
3.4012 V
2.4032 V
2.49345 V
1.49945 V
501.45 mV
14.76 mV
-1.00056 V
999.45 mV
-3.4363 V
-2.4343 V
-2.52456 V
-1.51856 V
-516.56 mV
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
3.4363 V
2.4343 V
2.52456 V
1.51856 V
516.56 mV
22.25 mV
-981.45 mV
1.01856 V
-3.4012 V
-2.4032 V
-2.49345 V
-1.49945 V
-501.45 mV
Ch1 10 mV Vert scale setting, 0 Div position setting
+3.5 V termination
+3.4 V offset
+2.4 V offset
1.5 V termination
2.5 V offset
+1.5 V offset
+0.5 V offset
V termination
0 V offset
-1 V offset
+1 V offset
V termination
-3.4 V offset
-2.4 V offset
-1.5 V termination
-2.5 V offset
-1.5 V offset
-0.5 V offset
3.4117 V
2.4137 V
2.4988 V
1.5195 V
521.5 mV
25.5 mV
-980.5 mV
1.0195 V
-3.4117 V
-2.4137 V
-2.5312 V
-1.5405 V
-538.5 mV
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
3.4483 V
2.4463 V
2.5312 V
1.5405 V
538.5 mV
34.5 mV
-959.5 mV
1.0405 V
-3.4483 V
-2.4463 V
-2.4988 V
-1.5195 V
-521.5 mV
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
122 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series