Commands Listed in Alphabetical Order
Examples
*PSC 0
sets the power-on status clear flag to false.
*PSC?
might return 1, indicating that the power-on status clear flag is set to true.
*RCL (No Query Form)
Restores the state of the power supply from a copy of its settings stored in the
setup memory. The settings are stored using the *SAV command. If the specified
setup memory is deleted, this command causes an error.
Group
Save and Recall
Syntax
*RCL <N R1>
Related Commands
*SAV
Arguments
<NR1> is an integer value in the range from 1 to 30 and specifies the location
of setup memory.
Examples
*RCL 3
sets the power supply to settings stored in memory location 3.
*RST (No Query Form)
This command resets the power supply to default settings, but does not purge
any stored settings.
Sending the *RST command does the following:
Returns the power supply settings to the defaults. (See page C-1, Default
Setup.)
Clears the pending operation flag and associated operations
Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual 2-23