Commands Listed in Alphabetical Order
*OPC
This command congures the instrument to generate an operation complete
message by setting bit 0 of the Standard Event Status Register (SESR) when all
pending comm
ands that generate an OPC message are complete.
The query command places the ASCII character "1" into the output queue when
all such OPC
commands are complete.
Group
Synchronization
Syntax
*OPC
*OPC?
Examples
*OPC? might return 1 to indicate that all pending OPC operations are nished.
*PSC
Sets and queries the power-on status ag that controls the automatic power-on
states of SRER and ESER. When *PSC is true, the Service Request Enable
Register (SRER) and Event Status Enable Register (ESER) are set to 0 at
power-on. When *PSC is false, the current values in the SRER and ESER
are preserved in nonvolatile memory when power is shut off and are restored
at power-on.
Group
Source
Syntax
*PSC <N R1>
*PSC?
Related Commands
*RST, *OPC
Arguments
<NR1> = 0 sets the power-on status clear ag to false, disables the power-on clear,
and allows the power supply to possibly assert SRQ after power on.
<NR1> 0 sets the power-on status clear ag to true. Sending *PSC 1 therefore
enables t he power-on status clear and prevents any SRQ assertion after power-on.
Returns
0|1
2-22 Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual