Performance Verication
DC Gain Accuracy
Performance checks Vertical scale Low limit Test result High limit
1 mV/div -1.5% 1.5%
2 mV/div -1.5% 1.5%
4.98 m V /div -3.0% 3.0%
5 mV/div -1.5% 1.5%
10 m V/div -1.5% 1.5%
20 m V/div -1.5% 1.5%
49.8 mV
-3.0% 3.0%
50 m V/div -1.5% 1.5%
100 m V /div -1.5% 1.5%
200 m V /div -1.5% 1.5%
500 m V /div -1.5% 1.5%
Channel 3 DC Gain
Accuracy, 0 V offset,
0Vverticalp
osition,
20 MHz BW, 1 M
1 V/div -1.5% 1.5%
1 mV/div -1.5% 1.5%
2 mV/div -1.5% 1.5%
4.9
8mV
-3.0% 3.0%
5 mV/div -1.5% 1.5%
10 m V/div -1.5% 1.5%
20 m V/div -1.5% 1.5%
4
9.8 mV
-3.0% 3.0%
50 m V/div -1.5% 1.5%
100 m V /div -1.5% 1.5%
200 m V /div -1.5% 1.5%
500 m V /div -1.5% 1.5%
Channel 4 DC Gain
Accuracy, 0 V offset,
0Vv
ertical position,
20 MHz BW, 50
1.0 V/div -1.5% 1.5%
1 mV/div -1.5% 1.5%
2 mV/div -1.5% 1.5%
4.98 m V /div -3.0% 3.0%
5 mV/div -1.5% 1.5%
10 m V/div -1.5% 1.5%
20 m V/div -1.5% 1.5%
49.8 mV
-3.0% 3.0%
50 m V/div -1.5% 1.5%
100 m V /div -1.5% 1.5%
200 m V /div -1.5% 1.5%
500 m V /div -1.5% 1.5%
Channel 4 DC Gain
Accuracy, 0 V offset,
0 V vertical position,
20 MHz BW, 1 M
1 V/div -1.5% 1.5%
MSO4000 and DPO4000 Series Specications and Performance Verication 25

Other manuals for Tektronix MSO2000 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO2000 series and is the answer not in the manual?

Tektronix MSO2000 series Specifications

General IconGeneral
BrandTektronix
ModelMSO2000 series
CategoryTest Equipment
LanguageEnglish

Related product manuals