Commands Listed in Alphabetical Order
TRIGger:A:BU
S:B<x>:SPI:DATa{:OUT|:MOSI}:VALue
Sets or returns the binary data string to be used for a SPI trigger if the trigger
condition is MOSI or MISOMOSI. Applies to bus <x>, where x is the bus number.
Conditions
This command requires a DPO2EMBD application module.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:SPI:DA Ta{:OUT|:MOSI}:VALue <QSt ring>
TRIGger:A:BUS:B<x>:SPI:DA Ta{:OUT|:MOSI}:VALue?
Arguments
<QString> is the binary data string w ith the number of bits specied by the
TRIGger:A:BUS:B<x>:SPI:DATa:SIZe command. The only allowed characters
in the
QString are0,1,andX.
TRIGger:A:BUS:B<x>:SPI:DATa:SIZe
Sets or returns the length of the data string to be used for a SPI trigger if the
trigger condition is MISO, MOSI, or MISOMOSI. Applies to bus <x>, where x
is the bus number.
Conditions
This command requires a DPO2EMBD application module.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:SPI:DA Ta:SIZe <NR1>
TRIGger:A:BUS:B<x>:SPI:DA Ta:SIZe?
Arguments
<NR1> is the length of the data string in bytes.
TRIGger:A:BUS:SOUrce
Sets or returns the source for a Serial bus trigger.
Conditions
This command requires a DPO2AUTO or DPO2EMBD application module.
Group
Trigger
2-312 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Other manuals for Tektronix MSO2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO2014 and is the answer not in the manual?

Tektronix MSO2014 Specifications

General IconGeneral
BrandTektronix
ModelMSO2014
CategoryTest Equipment
LanguageEnglish

Related product manuals