Installation
P6139A Passive Probe
Input Voltage:
400 V
RMS
or 400 V DC; CAT I (2,500 V
pk
transient)
300 V
RMS
or 300 V DC; CAT II (2,500 V
pk
transient
150 V
RMS
or 150 V DC; CAT III (2,500 V
pk
transient)
For steady-state, sinusoidal wave forms, derate at 20 dB/decade abo ve 2.5 MHz to 50 V
RMS
at 20 MHz and abo ve.
Output Voltage (terminated into 1 M Ω):
40 V
RMS
or 40 V DC; CAT I (2,5 00 V
pk
impulse)
30 V
RMS
or 30 V DC; CAT I (250 V
pk
impulse)
15 V
RMS
or 15 V DC; CAT I (250 V
pk
impulse)
Temperature:
Operating: -15 °C to +65 °C ( +5 ° F to +149 °F)
Nonoperating: -62 °C to +85 °C ( -80 °F to +185 °F)
Altitude: ≤ 2,000 meters
Pollution Degree: 2, Indoor use only
Humidity:
Operating: High : 40 °C to 50 °C, 10 % to 60 % RH
Operating: Low: 0 °C to 40 °C, 10 to 90% RH
MSO4000 and DPO4000 Series Oscilloscopes User Manual 11