EasyManuals Logo
Home>Tektronix>Test Equipment>MSO4032

Tektronix MSO4032 User Manual

Tektronix MSO4032
359 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #117 background imageLoading...
Page #117 background image
Acquire the Signal
3. If desired, push Undo Autos e t to undo the
last Autoset.
Undo
Autoset
Quick Tips
To position the waveform appropriately, Autoset may change the vertical position. Autoset always sets vertical offset t o 0 V.
If you use Autoset when no channels are displayed, th e instrument turns on channel one (1) and scales it.
If you use Autoset and the oscilloscope detects a video signal, th e oscilloscope automatically sets the trigger type to video and
makes other adjustments to display a stable video signal.
Acquisition Concepts
Before a signal can be displayed, it must pass through the input channel where it is scaled and digitized. Each ch annel has a
dedicated input amplier and digitizer. Each channe l produces a s tre am of digital data from which the instrument extracts waveform
records.
MSO4000 and DPO4000 Series Oscilloscopes User Manual 91

Table of Contents

Other manuals for Tektronix MSO4032

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO4032 and is the answer not in the manual?

Tektronix MSO4032 Specifications

General IconGeneral
BrandTektronix
ModelMSO4032
CategoryTest Equipment
LanguageEnglish

Related product manuals