Trigger Setup
9. If you are setting up a CAN trigger and have
made a Trigger On selection of Type of
Frame, push the lower-bezel Frame Type
button, and select Data Frame, Remote
Frame, Error Frame , or Overload Frame.
IfyouhavemadeaTrigger On selection of
Identier, push the lower-bezel Identier
button, and select a Format. Then push the
Identier side-bezel bu tton, and enter a binary
or hex value with multipurpose knobs a and b.
Push the lower-bezel menu Direction button
and select the direction of interest: Read,
Write,orRead o r Write.
If you have made a Trigger On selection
of Data. P ush the lower-bezel Data button
and enter the parameters of interest w ith
multipurpose knobs a and b.