Performance Verication
DC Offset Accuracy
All four-chan
nel models for channels 3 and 4, except as noted below
Performance checks Vertical scale Vertical offset
1
Low limit Test result
High limit
1 m V/div
900 m V 895.3 mV 9 04.7 mV
1 m V/div
-900 mV -904.7 mV - 895.3 mV
100 mV/div
5.0 V 4.965 V 5.035 V
Channel 3
DC Offset Accuracy,
20 MHz BW, 50
All four-channel 1 GHz
models
2
100 mV/div
-5.0 V -5.035 V -4.965 V
1 m V/div
900 m V 895.3 mV 9 04.7 mV
1 m V/div
-900 mV -904.7 mV - 895.3 mV
100 mV/div
9.0 V 8.935 V 9.065 V
100 mV/div
-9.0 V -9.065 V -8.935 V
500 mV/div
9.0 V 8.855 V 9.145 V
500 mV/div
- 9.0V -9.145 V -8.855 V
1.01 V/div
99.5 V 98.80 V 100.2 V
1.01 V/div
-99.5 V -100.2 V -98.80 V
5V/div
99.5 V 98.00 V 101.0 V
Channel 3
DC Offset Accuracy,
20 MHz BW, 1 M
All four-channel models
5V/div
-99.5 V -101.0 V -98.00 V
1 m V/div
900 m V 895.3 mV 9 04.7 mV
1 m V/div
-900 mV -904.7 mV - 895.3 mV
100 mV/div
5.0 V 4.965 V 5.035 V
Channel 4
DC Offset Accuracy,
20 MHz BW, 50
All four-channel 1 GHz
models
2
100 mV/div
-5.0 V -5.035 V -4.965 V
1 m V/div
900 m V 895.3 mV 9 04.7 mV
1 m V/div
-900 mV -904.7 mV - 895.3 mV
100 mV/div
9.0 V 8.935 V 9.065 V
100 mV/div
-9.0 V -9.065 V -8.935 V
500 mV/div
9.0 V 8.855 V 9.145 V
500 mV/div
- 9.0V -9.145 V -8.855 V
1.01 V/div
99.5 V 98.80 V 100.2 V
1.01 V/div
-99.5 V -100.2 V -98.80 V
5V/div
99.5 V 98.00 V 101.0 V
Channel 4
DC Offset Accuracy,
20 MHz BW, 1 M
All four-channel models
5V/div
-99.5 V -101.0 V -98.00 V
1
Use this value for both the calibrator output and the oscilloscope offset setting.
2
Due to system architecture differences between MSO/DPO4000B oscilloscope models, if the bandwidth of your model is less than 1 GHz, do not use the 50
term
ination setting on the oscilloscope to test offset accuracy. In order to tes t offset accuracy at the 50 termination setting for models less than 1 GHz, instead
usethe1M termination setting only. For 1 GHz models, test the DC offset accuracy at both the 50 termination setting and the 1 M termination setting.
Sample Rate and Delay Time Accuracy
Performance checks Low lim it Test result High limit
-1 divisions +1 divisions
38 MSO4000B and DPO4000B Series Specications and Performance Verication