Performance Verification
c. Set the Termina
tion (input impedance) to 1MΩ.
d. Repeat steps 9 through 13.
15. Repeat the pro cedure for all remaining channels as follows:
a. Push the front-panel button to deselect the channel that you have already
tested.
b. Move the DC voltage source connection to the next channel input to be
tested.
c. Starting from step 6, repeat the procedure for each channel.
MSO4000B and DPO4000B Series Specifications and Performance Verification 65