Performance Verication
Gain = V
bw-pp
/ V
in-pp
To pass the performance measurement test, Gain should be 0.707. Enter
Gain in the tes
t record.
13. Repeat steps 5 through 12 for all combinations of Vertical Scale and
Horizontal
Scale settings listed in the test record.
14. Repeat the procedure for all remaining channels as follows:
a. Push the front-panel button to deselect the c hannel that you have already
tested.
b. Move the calibrator connection to the next channel input to be tested.
c. Starting from step 3, repeat the procedure for each input channel.