EasyManua.ls Logo

Tektronix MSO70000C/DX

Tektronix MSO70000C/DX
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Verify the time base.
Equipment required Prerequisites
<4 GHz models: One precision 50 Ω coaxial
cable (Item 4)
<4 GHz models: One BNC to Minigrabber
adapter (item 18)
≥4 GHz models: One SMA cable (item 21)
≥4 GHz models: One adapter (item 19)
None
1. Initialize the instrument: Push the front-panel Default Setup button.
2. Hook up the signal source: Connect the probe compensation or fast edge
output to the Ch 1 input as shown in the following figure.
Figure 4: Setup for time base test
3. Set up the instrument: Push the front panel Autoset button.
4. Set the Vertical Scale to 200 mV /div (≤20 GHz models) or 120 mV/div
(>20 GHz models).
5. Set the time base: Set the horizontal Scale to 200 μs/div. The time-base
readout is displayed at the bottom of the graticule.
6. Verify that the time base operates: Confirm the following statements.
One period of the square-wave signal is about five horizontal divisions
on-screen for the 200 μs/div horizontal scale setting.
Rotating the horizontal Scale knob clockwise expands the waveform on-
screen (more horizontal divisions per waveform period),
counterclockwise rotation contracts it, and returning the horizontal scale
to 200 μs/div returns the period to about five divisions.
The horizontal Position knob positions the signal left and right on-screen
when rotated.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 107

Table of Contents

Other manuals for Tektronix MSO70000C/DX

Related product manuals