EasyManuals Logo

Tektronix MSO70000C/DX Technical Reference

Tektronix MSO70000C/DX
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
Sample acquisition mode,
typical
±(DC Gain Accuracy × | reading -(net offset) | + offset accuracy + 0.15 division + 0.6 mV)
Delta voltage measurement
between any two samples
acquired with the same setup
and ambient conditions,
typical
±(DC Gain Accuracy × | reading | + 0.15 division + 1.2 mV)
Convert offset, position and the constant offset term to volts by multiplying by the appropriate volts/
div.
Specification applies to any sample and to the High, Low, Max, Min, Mean, Cycle Mean, RMS, and
Cycle RMS measurements. Delta volts specification applies to subtractive calculations involving
two of these measurements. Delta volts specification applies to the Positive Overshoot, Negative
Overshoot, peak-peak, and amplitude measurements.
Position range ± 5 divisions
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Vertical system analog channels
(cont.)
6 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series

Table of Contents

Other manuals for Tektronix MSO70000C/DX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO70000C/DX and is the answer not in the manual?

Tektronix MSO70000C/DX Specifications

General IconGeneral
BrandTektronix
ModelMSO70000C/DX
CategoryTest Equipment
LanguageEnglish

Related product manuals