EasyManuals Logo

Tektronix TDS 224 User Manual

Tektronix TDS 224
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #46 background imageLoading...
Page #46 background image
Performance Verification
410
TDS 200 Series Digital Oscilloscope Service Manual
8. Change the instrument setup using the following steps:
Press menu button Select menu item Select setting
TRIGGER Slope Falling
9. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and
then check that triggering is stable.
10. Change the instrument setup using the following steps:
Press menu button Select menu item Select setting
TRIGGER Slope Rising
11. Disconnect the test setup.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 224 and is the answer not in the manual?

Tektronix TDS 224 Specifications

General IconGeneral
BrandTektronix
ModelTDS 224
CategoryTest Equipment
LanguageEnglish

Related product manuals