Warranted Characteristics
TDS 500B, TDS 600B & TDS 700A Performance Verification and Specifications
2–17
Table 2–10: Warranted Characteristics — Signal Acquisition System (Cont.)
Name Description
Delay Between Channels, Full
Bandwidth
TDS 600B: 100 ps for any two channels with equal Volts/Div and Coupling settings and
both channels’ deskew values set to 0
TDS 500B/700A: 50 ps for any two channels with equal Volts/Div and Coupling settings
Input Impedance, DC–1 MW Coupled 1 MW ±0.5% in parallel with 10 pF ±3 pF
Input Impedance, DC–50 W Coupled
50 W ±1% with VSWR 1.3:1 from DC – 500 MHz, 1.5:1 from 500 MHz – 1 GHz
Input Voltage, Maximum, DC–1 MW,
AC–1 MW, or GND Coupled
±300 V (DC + peak AC), 400 V peak; derate at 20 dB/decade above 1 MHz, CAT II
Input Voltage, Maximum, DC-50 W or
AC–50 W Coupled
5 V
RMS
, with peaks ±30 V
Lower Frequency Limit, AC Coupled 10 Hz when AC–1 MW Coupled; 200 kHz when AC–50 W Coupled
3
1
Net Offset = Offset – (Position × Volts/Div). Net Offset is the nominal voltage level at the oscilloscope input that
corresponds to the center of the A-D converters dynamic range. Offset Accuracy is the accuracy of this voltage level.
2
The limits given are for the ambient temperature range of 0_C to +30_C. Reduce the upper bandwidth frequencies by
5 MHz for the TDS 600B or by 2.5 MHz for the TDS 500B/700A for each _C above +30_C.
3
The AC Coupled Lower Frequency Limits are reduced by a factor of 10 when 10X passive probes are used.
Table 2–11: Warranted Characteristics — Time Base System
Name Description
Accuracy, Long Term Sample Rate and
Delay Time
TDS 600B: ±100 ppm over any 1 ms interval
TDS 500B/700A: ±25 ppm over any 1 ms interval

Other manuals for Tektronix TDS 500B