Warranted Characteristics
TDS 500B, TDS 600B & TDS 700A Performance Verification and Specifications
2–21
Table 2–14: Warranted Characteristics — Environmental (Cont.)
Name Description
Third Party Certification Conforms to and is certified where appropriate to:
UL 3111–1
3
CSA 22.2 no. 1010.1
3
1
VGA output cable needs to be terminated, if connected at all, for the Instrument to meet these standards. The test will
pass with LCOM part # CTL3VGAMM–5.
2
The GPIB cable connected to the instrument for certain of the emissions tests must be “low EMI” having a high-quality
outer shield connected through a low impedance to both connector housings. Acceptable cables are Tektronix part
numbers 012-0991-00, -01, -02, and -03. In order to maintain the EMI performance conforming to the above regulations,
the following cables, or their equivalent, should be used: a shielded Centronics cable, 3 meters in length, part number
012-1214-00, and a shielded RS-232 cable, 2.7 meters in length, CA part number 0294-9.
3
IEC 1010, UL 3111, CSA 1010 Safety Certification Compliance:
Temperature (operating) 5 to +40 C
Altitude (maximum operating): 200 meters
Equipment Type: Test and Measurement
Safety Class: Class I (as defined in IEC 1010–1, Annex H) – grounded product
Overvoltage Catregory: Overvoltage Category II (as defined in IEC 1010–1, Annex J)
Pollution Degree: Pollution Degree 2 (as defined in IEC 1010–1)
Note – Rated for indoor use only
Table 2–15: Certifications and compliances
EC Declaration of Conformity Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility and Low Voltage Directive
73/23/ECC for Product Safety. Compliance was demonstrated to the following specifications as
listed in the Official Journal of the European Communities:
EMC Directive 89/336/EEC:
EN 55011 Class A Radiated and Conducted Emissions
EN 55011 Class B Radiated and Conducted Emissions
EN 50081-1 Emissions:
EN 55022
Class B Radiated and Conducted Emissions
EN 60555-2 AC Power Line Harmonic Emissions
EN 50082-1 Immunity:
IEC 801-2 Electrostatic Discharge Immunity
IEC 801-3 RF Electromagnetic Field Immunity
IEC 801-4 Electrical Fast Transient/Burst Immunity
IEC 801-5 Power Line Surge Immunity
Low Voltage Directive 73/23/EEC
EN 61010-1 Safety requirements for electrical equipment for measurement,
control, and laboratory use
EN 61010-2-031:1994 Particular requirements for hand-held probe assemblies for electrical
measurement and test