Performance Tests
TDS 500B, TDS 600B & TDS 700A Performance Verification and Specifications
1–65
the other side of the BNC T connector. Now connect the BNC T
connector to CH 1. See Figure 1–24.
DC Calibrator
50 W Coaxial Cables
Dual Banana to
BNC Adapters
BNC T
Connector
Digitizing Oscilloscope
Figure 1–24: Subsequent Test Hookup
e. Measure amplitude of the probe compensation signal:
H Press SHIFT; then press ACQUIRE MENU. Press the side-menu
button AVERAGE then enter 16 using the keypad or the general
purpose knob.
H Adjust the output of the DC calibration generator until it precisely
overlaps the top (upper) level of the stored probe compensation
signal. (This value will be near 500 mV.)
H Record the setting of the DC generator.
H Adjust the output of the DC calibration generator until it precisely
overlaps the base (lower) level of the stored probe compensation
signal. (This value will be near zero volts.)
H Record the setting of the DC generator.
f. Press CLEAR MENU to remove the menus from the display. See
Figure 1–25.