Command Descriptions
2–134
TDS Family Oscilloscope Programmer Manual
NOTE. DCL does NOT clear the output queue once a hardcopy is in process. The
only way to abort the hardcopy process is to send the HARDCopy ABOrt
command. The output queue can then be cleared using DCL.
CLEARSpool clears the printer output spooler.
STARt initiates a screen copy that is sent to the controller where it can be stored
in a file or redirected to a printing device.
NOTE. Use the *WAI command between HARDCopy STARt commands to ensure
that the first hardcopy is complete before starting another.
HARDCOPY ABORT
stops any hardcopy output that is in process.
HARDCopy:FILEName
File System Only
Selects the file to send the hardcopy data to on the next hardcopy command
(HARDCOPY START). This is equivalent to setting the target file name in the
Hardcopy menu.
Hardcopy
HARDCopy
HARDCopy:FILEName <file path>
HARDCopy:FILEName?
<file path>
FILEName
:HARDCopy
?
<Space>
<file path> specifies that the hardcopy is sent to the named file. <file path>
is a quoted string that defines the file name and path. Input the file path using the
form <drive>/<dir>/<filename>.<drive> and one or more <dir>s are
optional. If you do not specify them, the TDS will write the file to the current
Examples
Group
Related Commands
Syntax

Table of Contents

Other manuals for Tektronix TDS 520B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 520B and is the answer not in the manual?

Tektronix TDS 520B Specifications

General IconGeneral
BrandTektronix
ModelTDS 520B
CategoryTest Equipment
LanguageEnglish

Related product manuals