Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–55
ALLOCATE:WAVEFORM:REF2 1000
reserves 1,000 data points for REF2.
ALLOCATE:WAVEFORM:REF1?
might return 500
APPMenu
Displays the user-definable Application menu, and the query returns the current
Application menu labels and title. This is equivalent to pressing the front-panel
APPLICATION button.
Application Menu
CLEARMenu, *ESR, EVENT?
APPMenu ACTivate
APPMenu?
APPMenu
<Space>
?
ACTivate
ACTivate displays the Application menu. Use the CLEARMenu command to
deactivate the Application menu.
Once the Application menu is activated, whenever a front-panel menu button is
pressed an event is generated that tells which button was pressed. See page 3–17
for event codes.
Menu button presses will also generate Service Requests when the URQ bit is
enabled in DESER and ESER and the ESB bit is enabled in SRER. See page 3–1
for a complete discussion of the use of these registers.
APPMENU ACTIVATE
displays the application menu.
Examples
Group
Related Commands
Syntax
Arguments
Examples

Table of Contents

Other manuals for Tektronix TDS 520B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 520B and is the answer not in the manual?

Tektronix TDS 520B Specifications

General IconGeneral
BrandTektronix
ModelTDS 520B
CategoryTest Equipment
LanguageEnglish

Related product manuals