Command Descriptions
2–64
TDS Family Oscilloscope Programmer Manual






 sets the specified channel to 50 W impedance.
 sets the specified channel to 1 MW impedance.
 
establishes 50 W impedance on channel 1.

might return , indicating that channel 3 is set to 1 MW impedance.
CH<x>:OFFSet
Sets or queries the offset, in volts, that is subtracted from the specified input
channel before it is acquired. The greater the offset, the lower on the display the
waveform appears. This is equivalent to setting Offset in the Vertical menu.
Vertical
CH<x>:POSition
 




 
 is the desired offset in volts. The range is dependent on the scale and the
probe attenuation factor. The offset ranges are shown below.
Arguments
Examples
Group
Related Commands
Syntax
Arguments