Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–179
defines the immediate measurement to be a frequency measurement.
MEASUrement:IMMed:UNIts? (Query Only)
Returns the units for the immediate measurement.
Measurement
MEASUrement:IMMed:TYPe
" !$$#
" !$
$#
$"! returns for volts, # for seconds, & for hertz, for
volts
2
,or for percent.
On the TDS 4XXA, $"! also returns for clocks, for volt-clocks,
or for 1/clks.
might return #, indicating that the units for the immediate measurement are
seconds.
MEASUrement:IMMed:VALue? (Query Only)
Immediately executes the immediate measurement specified by the MEASUre-
ment:IMMed:TYPe command. The measurement is taken on the source(s)
specified by a MEASUrement:IMMed:SOURCE command.
Measurement
" !$%
" !$
%
Examples
Group
Related Commands
Syntax
Returns
Examples
Group
Syntax