Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–41
ACQuire:NUMAVg, ACQuire:NUMENv, CURVe?, DATa:WIDth
For the TDS 400A, 510A, 500B & 700A:
ACQuire:MODe { SAMple | PEAKdetect | HIRes | AVErage |
ENVelope }
For the TDS 600B:
ACQuire:MODe { SAMple | PEAKdetect | AVErage | ENVelope }
For all TDS:
ACQuire:MODe?
For the TDS 400A, 510A, 500B, & 700A:
ENVelope
HIRes
ACQuire
:
MODe
<Space>
AVErage
PEAKdetect
SAMple
?
For the TDS 600B:
ENVelope
AVErage
ACQuire
:
MODe
<Space>
?
PEAKdetect
SAMple
SAMple specifies that the displayed data point value is simply the first sampled
value that was taken during the acquisition interval. In sample mode, all
waveform data has 8 bits of precision. You can request 16 bit data with a
CURVe? query, but the lower-order 8 bits of data will be zero. SAMple is the
default mode.
PEAKdetect specifies the display of the high-low range of the samples taken
from a single waveform acquisition. The high-low range is displayed as a
vertical column that extends from the highest to the lowest value sampled during
the acquisition interval. PEAKdetect mode can reveal the presence of aliasing or
narrow spikes.
Related Commands