Command Descriptions
2–76
TDS Family Oscilloscope Programmer Manual
CURSOR:PAIRED?
might return :CURSOR:PAIRED:UNITS BASE;POSITION1 -2.00E-3;POSITION2
2.00E-3;SELECT CURSOR2.
CURSor:PAIred:HDELTA (Query Only)
Queries the hbar (voltage) distance between the first and second paired cursor.
This is the absolute value of the vertical position of the first cursor minus the
vertical position of the second cursor.
Cursor
CURSor:FUNCtion
CURSor:PAIred:HDELTA?
CURSor
:
PAIred :
?
HDELTA
CURSOR:PAIRED:HDELTA?
might return 5.08E+0 for the voltage difference between the two cursors.
CURSor:PAIred:HPOS1 (Query Only)
Queries the horizontal bar (voltage) position of the first paired cursor.
Cursor
CURSor:FUNCtion
CURSor:PAIred:HPOS1?
CURSor
:
PAIred :
?
HPOS1
CURSOR:PAIRED:HPOS1?
might return -64.0E-3, indicating that the first cursor is at –64.0 mV.
Group
Related Commands
Syntax
Examples
Group
Related Commands
Syntax
Examples